What do I do if the 10AU Sensitivity FS% will not change?
When you are adjusting the sensitivity or basic operating level of the 10AU, if you can not observe any change in the FS% on screen 3.2 check and do the following:

•    Check to make sure the lamp and filters are installed properly.

•    Check to make sure you are in the manual gain control mode and in the proper range (High, Medium or low).

•    If after checking the above two items you are still experiencing problems go to screen 2.0 and press <6> reset calibration defaults to clear the internal settings and restore to factory defaults.

•    In extreme fouling conditions the FS % may not be adjustable.  The operator can check this when confirming the lamp/filters are installed correctly.