|
Technically Speaking,
It All Adds Up...
is a series of articles for people
who want to obtain the best possible results from their fluorometer.
Last
month Technically Speaking described how fluorometers can be
used for flow measurements using Rhodamine dye. This month's article
will discuss the importance of knowing where your readings fall
on the fluorophore response curve. It will summarize how to optimize
the accuracy of your results by ensuring you are in the most linear
part of that response curve.
Sometimes, fluorometer readings do
not change in step with the known concentration change of the fluorophore.
The factor that may be responsible for this is called "concentration
quenching", or sometimes just "quenching".
At low sample concentrations, the fluorescence
intensity is directly proportional to the sample concentration.
As the concentration increases beyond the linear range for fluorescence,
an effect known as "concentration quenching" occurs. At
these concentration levels, the sample starts absorbing so much
light that the light cannot pass through the sample for detection,
"concentration quenching" is occurring, see the representative
graph below.

Refer to the specifications of the
sensor you use for the maximum concentration that can be measured
in the linear region. For example, the new Cyclops sensor is specified
for a maximum concentration in the linear region of 1,000 ppb for
Rhodamine WT dye.
For best accuracy, unknown samples
should be tested for being in the linear part of the response curve
by noting the reading for the sample, then diluting the sample by
a factor of 10, and noting the new reading which should then be
a factor of 10 lower. The need to do this applies even if the initial
value appears to be in the linear range, see (1) below. There are
3 possible scenarios.
1. If the fluorometer reading increases when measuring the
diluted sample, then the concentration was high enough to be in
the sample quenching region, and the initial concentration value
was invalid. In this region, there is a reversal of fluorometer
readings, that is: diluting the concentration strength increases
the displayed reading. Dilute by a factor of 10:1 again, and note
if this time the reading decreases by a factor of 10:1. Repeat until
the sample concentration is in the linear range.
2. If the fluorometer reading decreases, but not by a factor
of 10, then the concentration level is outside the linear region,
but could be accurately read by generating a "calibration curve"
for the sample/test configuration and applying a correction factor.
3. If the fluorometer reading decreases by a factor of 10,
then the initial sample concentration was within the linear range
and was accurately read.
For (1) and (2) above, the final fluorometer
reading should be multiplied by the dilution factor to obtain the
absolute sample concentration.
|