Chapter
1:
Fluorescence Theory
Chapter
2:
Advantages of Fluorescence
Chapter
3:
Instrumentation
Chapter
4:
Variables of Fluorescence
Chapter
5:
Calibration and Standards
|
An
Introduction to
Fluorescence
Measurements
Chapter 5
- Calibration and Standards
5.1 Calibration of a Filter Fluorometer
Filter fluorometers can be calibrated
by a number of different techniques. The most common calibration of
a filter fluorometer consists of compensating for blank (solution containing
zero concentration of the substance to be read) and adjusting the instrument
to reflect a known concentration of sample (the standard). Filter fluorometers
can also be calibrated by correlating their raw fluorescence signal
compared to a standard laboratory method like EPA method 413.1.
Fluorescence is a relative measurement
and the optics and electronics of each instrument vary, certainly from
manufacturer to manufacturer, but also among instruments from the same
manufacturer. A fluorometer must be calibrated and recalibrated whenever
the optics or filters are changed.
Fluorescence is subject to temperature
and other environmental effects; it is important to calibrate the fluorometer
in conditions as close as possible to the actual conditions for your
study. Recall that fluorescence is an extremely sensitive measurement.
Sample readings are only as accurate as the standard and blank used
to calibrate the instrument. It is important to be rigorous in laboratory
procedures, such as cleaning labware and carefully preparing standards.
Instrument stability is critical to accurate
readings. If frequent calibration is impractical or for long-term studies
or on-line monitoring, a fluorometer with stable reference circuitry
(closed-loop) like the Turner Designs 10-AU or TD-4100 should be used.
See section 3.53.
5.2 Standards
5.21 Standards.
A blank should be obtained before measuring sample. The blank is water
or matrix liquid for the sample, taken before any of the substance to
be measured has been added. This liquid should be used as the make-up
liquid for standards and to set the instruments reading to zero.
Normally, the standard will be a known
concentration of the material to be quantitated. For single-point calibrations,
and where a known concentration is needed, it is best to choose a standard
with a concentration approximately 80% of the highest concentration
to be read. A less concentrated standard may be used, but a higher concentration
will provide greater accuracy.
If concentrations above the linear range
are to be read, several standards should be used so a calibration curve
may be prepared. For example, if the substance is linear to 250 ppm
and measurable with a calibration curve to 1000 ppm, you might calibrate
with a 200 ppm standard and take readings at 500, 750, and 1000 ppm
for a calibration curve.
In some cases, calibration with a standard
of known concentration is not necessary. In procedures such as in
vivo chlorophyll or certain flow measurements, instrument sensitivity
will be set with an unknown concentration. For in vivo chlorophyll
measurements, for example, in most cases the "standard" will
be an unknown sample from the body of water under investigation. During
the sensitivity-setting procedure, while the unknown is running through
the fluorometer, you should take a grab sample of the unknown immediately
after it passes through the flow cell. This grab sample will be extracted
later and the actual chlorophyll concentration determined. The researcher
will then use a ratio calculation method to compare in vivo sample
readings with extracted readings to determine the actual concentration
of field samples.
Where the instrument has more than one
range or "door" factor, many researchers will want to check
for range-to-range correlation. Some fluorometers, such as the Turner
Designs 10-AU are designed with excellent range-to-range correlation,
making this kind of comparison unnecessary in most cases.
The stability of the standard is an important
issue for accurate readings. Some substances, such as Rhodamine WT are
stable for months. Others, such as chlorophyll will become inaccurate
in as little as a few hours when exposed to light or temperature changes.
Properly store standards to ensure stability or make new ones just prior
to calibration. In some cases, purchasing premade standards is a practical
and economical option. Contact Turner Designs for information about
extracted chlorophyll and Rhodamine WT standards.
5.22 Secondary Standards.
The stability and accuracy of standards is a critical factor in fluorescence
measurements. In cases such as chlorophyll measurements, where standards
are not stable over the long term, a stable secondary standard can considerably
simplify procedures, is relatively inexpensive, and can provide confidence
in the accuracy of readings.
A secondary standard is basically a substance
that fluoresces at wavelengths similar to the substance being measured.
Turner Designs, for example, offers secondary standards for many applications
that are stable for years with no special storage requirements. Secondary
standards are used in place of the primary standard after the instrument
has been calibrated at least once with the primary standard.
For example, if measuring chlorophyll,
the researcher would first obtain or prepare accurate chlorophyll standards
and calibrate with these primary standards. (Premade, ready-to-use chlorophyll
standards in 90% acetone are available from companies such as Turner
Designs. Chlorophyll a in a dry powder ready for diluting in
90% acetone or methanol is available from chemical companies such as
Sigma or Fluka.) The researcher would then read the secondary standard
and record the concentration as compared to the primary standard. After
this, the secondary standard could be used in place of the primary standard
when calibrating.
Another advantage to a stable secondary
standard is as a check on instrument stability. A secondary standard
can be read periodically, and if the reading has not changed from the
last calibration, the researcher can proceed with confidence that the
instrument is providing accurate readings. If the reading has changed
significantly, simply recalibrate using the secondary standard. A secondary
standard is especially useful for instruments that lack reference circuitry.
A secondary standard greatly minimizes the need to constantly prepare
fresh standards, which is expensive and time-consuming, as well as potentially
introducing error from contaminated or improperly prepared standards.
1. Lakowicz, J.R. 1983. Principles
of Fluorescence Spectroscopy, Plenum Press, New York.
2. Guilbault, G.G. 1990. Practical
Fluorescence, Second Edition, Marcel Dekker, Inc., New York.
3. Id., p. 7.
4. Dr. Richard Thompson. 1998. University
of Maryland, Department of Biochemistry and Molecular Biology, School
of Medicine.
5. G. K. Turner, "Measurement of
Light From Chemical or Biochemical Reactions," in Bioluminescence
and Chemiluminescence: Instruments and Applications, Vol. I, K.
Van Dyke, Ed. (CRC Press, Boca Raton, FL, 1985), pp. 45-47.
6. Guilbault, G.G. 1990. Practical
Fluorescence, Second Edition, Marcel Dekker, Inc., New York, pp.
51-57.
7. Lakowicz, J.R. 1983. Principles
of Fluorescence Spectroscopy, Plenum Press, New York, chap. 2.
8. Guilbault, G.G. 1990. Practical
Fluorescence, Second Edition, Marcel Dekker, Inc., New York, pp.
67-69.
9. Lakowicz, J.R. 1983. Principles
of Fluorescence Spectroscopy, Plenum Press, New York, pp. 23-26.
10. Guilbault, G.G. 1990. Practical
Fluorescence, Second Edition, Marcel Dekker, Inc., New York, pp.
57-58.
11. Stotlar, S. C. 1997. The Photonics
Design and Applications Handbook, 43rd Edition, Laurin Publishing
Co., Inc., Pittsfield, MA, p. 119.
12. Kodak Filters for Scientific and
Technical Uses, Eastman Kodak Company, 3 ed. 1981.
13. Andover Corporation Optical Filter
Guide, Andover Corporation.
14. Id.
15. Kodak Filters for Scientific and
Technical Uses, Eastman Kodak Company, 3 ed. 1981.
16. Guilbault, G.G. 1990. Practical
Fluorescence, Second Edition, Marcel Dekker, Inc., New York, p.
63.
17. Dr. Richard Thompson. 1998. University
of Maryland, Department of Biochemistry and Molecular Biology, School
of Medicine.
18. Guilbault, G.G. 1990. Practical
Fluorescence, Second Edition, Marcel Dekker, Inc., New York, p.
30.
19. Dr. Richard Thompson. 1998. University
of Maryland, Department of Biochemistry and Molecular Biology, School
of Medicine.
20. Iain Johnson, Product Manager, and
Ian Clements, Technical Assistant Specialist (May 1998 communication
from Molecular Probes, Eugene, Oregon).
21. Fluorometric Facts: A Practical
Guide to Flow Measurement, Turner Designs (1990), pp. 14-15.
22. Guilbault, G.G. 1990. Practical
Fluorescence, Second Edition, Marcel Dekker, Inc., New York, p.
172.
23. Fluorometric Facts: A Practical
Guide to Flow Measurement, Turner Designs (1990), p. 21.
24. Guilbault, G.G. 1990. Practical
Fluorescence, Second Edition, Marcel Dekker, Inc., New York., p.
28.
Continue...
to beginning of paper
|