Primary Chlorophyll Standards
Liquid Primary Chlorophyll
a Standards (Part No. 10-850)
Preparing primary chlorophyll a standards
for fluorometry can be challenging, as well as time consuming. Traditionally,
pure chlorophyll a is purchased from a manufacturer or purified
via HPLC. Then, an initial dilution of this pure extract is measured using
a spectrophotometer to determine the exact concentration, before diluting
to the range for a fluorometer. Stability of the final standard is also
a concern, since it is temperature sensitive as well as photosensitive.
Turner Designs offers a solution to this challenging and time consuming
process.
Features of the Liquid Primary Chlorophyll
Standards
- Saves Time and Money. Spectrophotometrically
determining the chlorophyll concentration is capital and labor intensive.
The concentration of the liquid standards offered by Turner Designs
has already been determined with a spectrophotometer.
- Simple and Accurate. Turner Designs
liquid chlorophyll standards are sealed in an ampule and ready to use.
This eliminates the need to purchase, measure, or prepare raw chlorophyll
for standards. Measurements are simplified and accuracy is improved.
- Guaranteed. If stored properly,
the liquid chlorophyll standards are guaranteed for two months after
arrival in your lab.*
- Documented. Each set of liquid
standards comes with a certified concentration and certificate of analysis
for reference.
Turner Designs Offers Solutions
Chlorophyll a prepared in 90% acetone/10%
distilled water for extractive analysis is now available from Turner Designs.
The set of two ready-to-use liquid fluorometric standards consists of
one high-level and one low-level concentration. Standards are essentially
free from chlorophyll b and are diluted with solvents that are
free of contaminants that can cause background fluorescence.
*Standards must be stored at -20° C in their
original foil wrap and opened under subdued lighting conditions.
Three Simple Steps to Calibrating:
- Break the ampule.
- Pour the contents of the ampule into a
cuvette.
- Calibrate your fluorometer.
Chlorophyll a Concentration Specifications:
Parent Standard (kept by Turner Designs for
Q.C.): 1 mg/L. Concentration determined using a spectrophotometer and
the Lorenzen monochromatic equations. Standard purity measured through
HPLC analysis.
High Fluorometric Standard: 20 ml of approximately
155 µg/L
Low Fluorometric Standard: 20 ml of approximately
15.5µg/L
The concentration of each lot will vary,
but the final concentration will be listed on the certificate of analysis
and reported to three significant figures.
|