
Learn
more about the 10-AU Fluorometer
Solid
Secondary Rhodamine WT Standards now available from Turner Designs!
|
Flow Measurements in Sanitary
Sewers
by Dye Dilution
Foreword
Flow
in sanitary sewers may be measured to an accuracy of about ±1%, using
fluorescent dye dilution techniques. It is the most cost- effective method
for:
- Quantitating
and localizing infiltration.
- Calibrating
any type of existing flow meter.
This
monograph, based on extensive experience in the United States and in Canada,
is basically a "how to do it" manual. It complements the more
theoretical monograph entitled "A PRACTICAL GUIDE TO FLOW MEASUREMENT",
available at no cost from TURNER DESIGNS.
- It
works in surcharged manholes.
- There
is no need to crawl a manhole.
Because
this monograph will anticipate unusual situations, as well as discuss
normal situations, the procedure might at first glance seem complicated.
In practice, however, in most instances it is straightforward. Since there
is no need to crawl a manhole, the procedure can be done by one person.
An experienced person can carry out the entire procedure, from preparation
of dye to calculation of flow in one manhole, in about 25 minutes. Flows
at subsequent manholes downstream simply involve popping the manhole cover,
dropping a can on a string to collect a sample, and a few minutes for
the reading of the sample and the calculation. Samples may be measured
on the spot, or if preferred, may be taken to the office or a lab and
measured at your convenience.
Definitions
Fluorescent
Materials. A fluorescent material is one that absorbs light of one
color and emits light of a different color. For practical purposes, the
emitted light is always of longer wavelength, i.e., shifted toward the
red. The fluorescent tracer material under discussion, rhodamine WT, absorbs
green light and emits red light. By contrast, brighteners, which are fluorescent
materials added to detergents, absorb ultraviolet light and emit blue
light.
Sewage
Blank. The blank is amount of fluorescent materials occurring naturally
in the system, expressed as a concentration of the fluorescent material
being measured. The great value of rhodamine WT lies in the fact that
the wavelengths involved are unusually long. A high concentration of brightener
looks like distilled water to the instrument when it is set up for measuring
rhodamine WT. A few naturally occurring materials and some industrial
materials will fluoresce very weakly at the wavelength of rhodamine WT.
The highest "blank" we have encountered in sanitary sewage was
equivalent to about 2 parts per billion of the 20% solution of rhodamine
WT. Most ranged from 0.4 - 1.5 parts per billion.
It
should be noted that fluctuation of blank during a measurement, not the
absolute magnitude, is the important factor. (See DYE INJECTION -- Dye
Concentration.)
Sewage
Standard. A "standard" is a known concentration of the material
of interest. In this case, the standard will be a known dilution of the
dye being injected. The dilution will be made with sewage collected just
prior to the start of dye injection.
Recovery
Standard. We will also refer to a "recovery standard", which
will be made by diluting the dye being injected with deionized water (distilled
water). The purpose of the "recovery standard" is to see whether
the reading of a given concentration of dye is the same in both clean
water and sewage -- in other words, whether the dye is all "recovered"
in sewage.
Dilution
Ratio. This is the entire basis of the technique. If you inject dye
at a constant rate of 0.01 gpm and find the sewage dilutes it one-million
fold, then the dilution ratio is one million. The flow rate in the sewer
must then be one million times as great as the injection rate, or 10,000
gpm. Note that it is not necessary to know the absolute concentration
of the dye in either the solution injected or in the sewer -- only the
dilution ratio. The instrument will be calibrated with a known dilution
of the injected dye solution.
Procedure
Outline
- Collect
about one gallon of sewage immediately prior to injecting dye. This
will be used for preparing standards.
- Add
dye to the sewage system at a constant rate. This will be discussed
under DYE INJECTION.
- Take
samples downstream of the point where the dye is added. This will be
discussed under SAMPLING.
- On
rare occasions, a very quick filtration of the sample will be required.
Elapsed time will be about five minutes, and many samples may be filtered
at the same time. The filtration equipment is so inexpensive that it
is usually discarded. This will be discussed under FILTRATION.
- Measure
the dilution ratio of the dye in your samples. This will be discussed
under PREPARATION OF STANDARDS and MEASUREMENT.
- Calculate
the flow. This will be discussed under FLOW RATE CALCULATION.
- Sewage
is not all the same -- so, you should run a recovery ratio occasionally.
This will tell you whether sample processing is really necessary and
how effective the processing is. This is discussed under PROOF OF PERFORMANCE.
Equipment
Needed
For
flow measurements in sanitary sewers, if you are going to be using only
the grab sample method in an area which is reasonably protected from the
weather, the laboratory fluorometer will be sufficient. If, however, you
want to use the grab sample method in less favorable weather or if you
wish to use a continuous-flow method, you will need the field fluorometer.
If
you are also interested in chlorophyll, nephelometry (turbidity), circulation,
effluent dispersion, mixing, or outfall location, you may wish to contact
Turner Designs for the best selection of equipment to meet your overall
requirements.
Fluorometer
(select one)
- 10-AU-000
Laboratory Fluorometer (CANNOT perform continuous-flow studies)
- Sample
system: 25 mm x 13 mm Discrete Sample Holder Set
- Filters:
10-041R Rhodamine Optical Kit
- 10-AU-005
Field Fluorometer
- Sample
system: 25 mm x 13 mm Discrete Sample Holder Set for grab samples; OR
the 10 mm
- Continuous-flow
cuvette system for flow-through measurements
- Filters:
10-041 Rhodamine Optical Kit
Case
(select one)
- 10-AU-002
Indoor Case
- 10-AU-022
Watertight Case
Cuvettes
10-031,
the 13 mm x 100 mm selected cuvettes (for grab samples)
Other
equipment needs are discussed in the sections entitled DYE INJECTION,
FILTRATION and PREPARATION OF STANDARDS.
Dye
Injection
Successful
dye injection involves several factors. Injection must be at a constant
concentration, at a constant, precisely known rate, and without surging
or splashing.
Dye
Source and Characteristics
Rhodamine
WT may be ordered from Crompton & Knowles Corp., P.O. Box 33157, Charlotte,
NC 28233-3157. Telephone: 1- 800-323-4383. Turner Designs also supplies
the dye in manageable quantities (Part Number 10-108). Contact Turner
Designs for a more complete list of dye suppliers.
Rhodamine
WT is supplied as a clear, 20% aqueous solution. It is very dark red in
color, somewhat viscous, and has a specific gravity of 1.2. The concentrated
dye is extremely irritating to the eyes.
It
was specifically developed to have poor "dye" properties, in
the conventional sense. This means that is does not absorb significantly
on sediments, which would reduce its value as a tracer. As a bonus, it
may be removed from your hands with soap and water, though it is not easily
removed from clothes.
Dye
Concentration
Throughout
this manual, all guideline concentrations are based on considering the
material as purchased to have a concentration of 100% on a basis of volume.
The
fact that the material is supplied as a 20% aqueous solution, with a specific
gravity of 1.2 is not significant. It needs to be mentioned only because
much of the technical literature considers dye concentration in terms
of active ingredient.
The
concentration of the dye to be injected is governed by three things: injection
rate; estimated flow in the system; and final concentration desired.
As
long as the dye injection rate is small compared to the flow being measured,
it may be independently specified. Usually, it will be determined by the
injection system which you purchase or build.
We
find that injection rates much below 10 ml/minute (0.003 gpm) are difficult
to handle.
The
size of the dye reservoir may set the upper limit of injection rate. If,
for example, you are doing infiltration studies on sewer systems, where
dye injection periods may be several hours long, convenience in handling
the reservoir may limit the maximum injection rate.
Dye
concentration in the final sample taken should be high enough so that
variability in the blank -- see DEFINITIONS -- will not affect accuracy.
It should be low enough so that readings are linearly related to dye concentration.
With
the 13 mm x 100 mm Cuvette Holder recommended, this final dye concentration
should fall between 0.02 and 0.5 parts per million (20 - 500 ppb), for
most work.
In
extremely large systems, or where multiple measurements are to be made,
dye costs may make it worthwhile to explore blank variability, and aim
for a lower final dye concentration.
Take
perhaps 10 blank samples over a one hour period and determine the blank
variability. If you need only 5% accuracy, and dye quantity is a factor,
then aim for a concentration 20 times the variability of the blank.
In
situations where injection will be at one point and sampling at multiple
points (as in infiltration studies), the flow change may be such that
you must exceed 0.5 ppm at the early sample points. Up to 4 ppm can be
measured with a multi-point calibration curve. See MEASUREMENT - High
Concentration Measurements. Any concentration can be measured by dilution.
Simply make a 10-fold or 100-fold dilution and multiply the answer by
10 or 100. You will probably dilute with distilled water and must remember
that your blank and standards should be diluted with distilled water in
the same ratio.
So,
decide on your desired final dye concentration and estimate the flow to
be measured. Units, of course, do not matter if they are consistent. Call
your estimated flow Q. The flow-rate of your injector is q, in the same
units. Assuming you desire a final dye concentration of 0.1 ppm, the injected
dye concentration should be:
0.1 Q/q,
in ppm
For
example, if your estimated flow-rate (Q) is 1000 gpm and your injector
delivers .01 gpm (q), your initial dye concentration should be about 10,000
ppm, or 1%. Add 1 part of dye to 100 parts of water.
Two
simple guidelines:
- We
recommend the use of an RB-BG75- 2CSY pump for dye injection (see below).
When set for maximum flow, it delivers 48 ml/min., which translates
into about 2 x 10-5 million gallons per day (mgd).
With this injection rate, we recommend that you estimate the
flow in mgd, then multiply by 0.005 to get the desired initial dye
concentration. For 0.1 mgd, use 0.05% dye; for 1 mgd, use 0.5% dye;
for 10 mgd, use 5% dye, etc.
This
will yield a final concentration of about 0.1 ppm which is normally
a good concentration to shoot for. If your actual flow is one-fifth
your estimate, dye concentration will still be in the linear range.
If flow is five times higher than estimated, the concentration will
still be about 20 times the background normally encountered. This
is conservatively high as a 20% change in background during the run
would result in only a 1% error.
- The
concentration of the dye being injected does not need to be accurate!
As
an example, the scintillation vials which we recommend for sample
storage (see Sample and Standard Storage, below) contain about 2/3
ounce, when filled to the bottom of the shoulder. This translates
to about 0.005 gallons.
If
you have estimated flow to be one mgd, and hence need a 0.5% dye concentration,
it is easy to prepare. Fill a scintillation vial to the bottom of
the shoulder with concentrated dye. Pour this into a plastic one-gallon
bottle. Mix and fill with water, to obtain the desired 0.5% dye concentration.
With
the pump recommended, the gallon will last about 1.3 hours.
Mixing
the Dye
In
mixing the dye, there are several things to keep in mind.
- If
it is not uniformly mixed, the results will be inaccurate!
- A
completely full container is impossible to mix. Leave at least 20% or
more air space.
- A
container you can pick up and shake is highly recommended.
- The
dye acts like a detergent; it foams. Never try to add more water or
dye after mixing, or you will have red foam running down the sides.
- If
the container can be picked up, mix by swirling the solution inside
using a rotary motion, then turning it upside down. Swirl again and
right it. Do this 10 times.
For
large drums, take a smaller container and use part of the water to
pre-dilute the dye 20 - 50 fold. Put the rest of the water required
in the drum. After the foam settles in the pre-diluted dye, pour it
into the drum. Seal it, and if you can't find someone with a drum
shaker, you must roll the drum repeatedly around on the floor, occasionally
setting the drum on end to change the flow pattern. You can establish
how much mixing is required by taking samples from several levels,
diluting them and reading them in the fluorometer.
Injection
Apparatus
The
most convenient device is a constant displacement pump. A commercially
available pump which appears to be nearly ideal for field work is the
model RP-BG75-2CSY pump made by Fluid Metering, Inc., and available through
Turner Designs. This unit features continuously variable flow rate and
is self- priming. Maximum flow rate is 48 ml/min. (0.012 gallons per minute)
(2 x 10-5 mgd). Current drain is between 0.06 and 0.1 amperes from a twelve
volt battery. A very light duty storage battery, or a cigarette lighter
adaptor for your vehicle may be used. Eight size "D" alkaline
flashlight batteries (10 ampere-hour at 1.5 volts per cell) should run
the pump for about 100 hours. The pump which we have used for field demonstration
has shown an excellent reproducibility of ±0.1% (at maximum rate) over
a period of 5 years.
CAUTION!
These pumps are not explosion proof. The dangers of sewer gas explosion
must be taken into account.
A
modified chlorine feeder, available from Aerofeed, Inc. (P. O. Box 3203,
Chalfont, PA, 18914), may also be used. It is pressurized in the field
with a foot or tire pump. The flow rate is adjustable from 60 to 250 ml/min.
(.016 to .066 gallons per minute). It features a self-contained reservoir
with a 9 liter (2.4 gallons) capacity.
The
flow is controlled by a special pressure regulator, which maintains a
constant pressure difference of 5 pounds per square inch cross a metering
valve.
NOTE:
We know these devices have been used successfully, but have not tested
them in our laboratory.
A
device frequently mentioned in the literature, the Mariotte Vessel is
described in a companion monograph entitled "A Practical Guide to
Flow Measurement." While much used in stream gaging, it is rarely
used in sewer studies. Its major advantage, the ability to deliver large
injection rate without power, is not important in sewers. There are also
numerous precautions and disadvantages (see "A Practical Guide to
Flow Measurement"), not the least of which is that one small enough
to be handled easily by one person is tricky to build and subject to plugging.
Things
We've Learned the Hard Way!
- Dye
should not be allowed to free fall for more than about one or two feet,
or splashing and erratic results may occur.
- If
you are using a small injection system, place reservoir, pump, and battery
in a box with a yoke on it, and suspend it by rope from street level.
Tie a knot in the rope, and you can close the manhole for the duration
of the test.
- Do
not attempt to use a long tube, at low flow rates from street level.
The dye flow will surge, and the erratic suction on the pump exhaust
line will cause erratic results.
- In
fact, long tubes are to be discouraged from another standpoint. Flexible
vinyl tubing (e.g., Tygon©) is convenient, but it does adsorb dye reversibly.
It does have a limited capacity for adsorbing dye and at the concentrations
injected, equilibrium is rapidly achieved. A few feet is no problem.
A hundred feet is unacceptable. Polyethylene, polypropylene, and unplasticized
vinyl tubing do not adsorb, but are fairly rigid.
- If
you must inject from street level, use an appropriate length of PVC
pipe with a tee on the top. Insert the tube from your injection pump
well down the pipe or allow the stream from your dye injector to fall
in the top, clearly missing the wall until well down. Run water from
a hose through the tee. This is a particularly valuable technique for
injecting dye through a pump directly into the bell housing of a pump,
or for other situations where the desired point of injection is not
directly below an opening. The PVC pipe is flexible and easily maneuvered.
The top is easily secured to a ladder laid across the manhole, etc.
- If
the amount of water added in injecting the dye is substantial compared
to the flow being measured, consider correcting for the added water.
In infiltration studies, it will normally be a constant increase in
baseline, and therefore can be ignored. In calibrating a flow meter,
the added flow should be subtracted from the measured flow for comparison
with an upstream flow meter. IF the flow meter is downstream, there
will be no correction, as the flow meter is reading the total flow.
Where
Should the Dye be Injected?
The
dye injection and dye sampling points must be considered together.
The
important considerations are:
- The
dye must be injected sufficiently far from the sampling point so that
it is completely mixed with the sewage before samples are taken.
This may easily be checked. See SAMPLING -- Is
Mixing Adequate?
The
more turbulent the flow, the better. A pass through a lift pump virtually
assures mixing.
Even
under relatively smooth flow conditions, expect good mixing 100 to
200 pipe diameters down stream.
Avoid
situations where there is a reservoir between the injection and sampling
points. It will increase the time required to make a test.
- If
flow increases downstream, dye dilution techniques measure the total
flow at the point where the sample is taken!
If there is infiltration or a branch entering
between the injection and sampling point, you will be measuring the
total flow at the point where the samples are taken.
Exfiltration
is a special situation. Assuming complete mixing prior to loss of
flow, the flow measured is that prior to the diversion. To quantitate
such situations requires moving the injection site as well as the
sample site.
- Convenience
should be considered, particularly at the sampling point. Certainly,
the first time around you will wish to take a fair number of samples.
Calibrating
the Injector
The
accuracy of your final result depends directly on the accuracy of your
knowledge of the rate of flow of the dye.
You
will need a stopwatch and a clean, dry volumetric flask whose capacity
will depend on the type of injection system you are using. See PREPARATION
OF STANDARDS for information on volumetric flasks.
When
dye flow has stabilized, measure the time to fill the volumetric flask,
then calculate the flow rate. See FLOW RATE CALCULATIONS.
We
strongly recommend a flow rate check both before and after a run.
When
using the Fluid Metering, Inc., pumps, a volumetric flask should be chosen
that takes at least four minutes to fill. This is because the pump makes
only 75 strokes per minute. (It is also the reason that we do not recommend
the RP-BG 25 series pumps. They make only 25 strokes per minute and it
takes too long to accurately calibrate them.)
Save
the dye collected during calibration of the dye injector. See SAMPLING.
Sampling
What
Samples are Needed?
Collect
the following:
- A
sample of the dye being injected. This was collected when the dye injector
was calibrated.
This
dye sample will be used to prepare standards.
- A
large sample of sewage (1 gallon) to which dye has not been added.
This
will be used to determine BLANK (see DEFINITIONS, above).
It
will also be used to dilute the dye sample mentioned in 1, above,
when preparing the calibration standards.
- Samples
of sewage taken downstream of the injection point. One ounce will be
adequate.
- You
may wish to take additional samples upstream of the dye injection point,
to check for variability in the BLANK. See 2, above. One ounce will
be adequate.
When
Should Samples Be Taken?
If
you were to plot dye concentration as a function of time from the start
of dye injection, you would find:
- A
delay time. This delay time is almost invariably longer than you would
expect from an estimate of velocity. It is not unusual to find a delay
time of 15 - 20 minutes for dye to travel 500 feet.
- A
period during which the dye concentration rises. If the flow is highly
turbulent, this period will be short. If flow is smooth, and particularly
if there is a reservoir of any kind between the dye injection point
and the sampling point, this period may be quite long.
- A
period of constant dye concentration (plateau) assuming that flow rate
is constant.
- A
fall off in concentration of about the same duration as the rise time,
after the injection is discontinued.
The time periods involved are a function of travel
time between injection and sample sites. In a reach of 100 yards in
a fast moving stream, a 15-minute injection might show a rise time
of 1-2 minutes, a plateau of 11-12 minutes, and a fall off time of
1-2 minutes. A mile downstream, there might be no plateau at all without
longer injection.
A
major advantage of the Model 10-AU Fluorometer is that you can take it
into the field and power it from your car battery. Cigarette adapters,
available at any auto parts store, will give you instant power.
By
using the fluorometer in the field, you can be sure of results. Take samples
at regular intervals, typically one every two minutes, and measure them
on site until readings stabilize. These readings may not be used for final
flow measurements, but they will enable you to know when to take the sample
for flow calculations.
When
the readings are stable, you may take the one-ounce samples required for
final flow determination.
Is
Mixing Adequate?
When
readings become constant as a function of time, as described immediately
above, take samples across the stream flow. If these samples read the
same, mixing is adequate.
How
Many Samples are Needed?
Theoretically,
only one sample is needed, taken any time during the plateau period. It
is good insurance to take at least three at 1 minute intervals. It is
not at all unusual to accidentally contaminate a sample. If you have only
two samples, and they disagree you won't know which is right. If you have
three or more samples and one is wild, you can discard it with an easy
mind.
For
very accurate calibration of a flow meter, you may wish to take 10 samples
and average the results.
Sample
and Standard Storage
Your
samples and standards are very stable and may be kept for months for recheck
if desired. For storage longer than one day, they should be protected
from strong sunlight.
The
storage container material is important. A tin can on a rope is fine for
collecting, but the samples should not be stored in metal. Also, some
types of ordinary glass jars degrade the sample. This degradation may
be a pH effect due to traces of lime left on the surface of the glass.
Rubber and some plastic liners will absorb the dye. Potential containers
should be tested by filling several with a solution of dye (about 0.1
- 0.3 ppm). If the samples are likely to be jostled in transport, turn
the containers upside down so the lid contact and seal will be tested.
A few hours later, compare dye concentration with the original material
stored in your volumetric flask.
Materials
known to be safe are:
1.
Borosilicate glass, such as Pyrex® or Kimax®.
2.
Polyethylene, polypropylene, and polystyrene.
3.
Polyseal® screw caps (available for narrow mouth bottles). Note: If you
have trouble with lids, but not the bottles, you can always screw the
lid on over a piece of Parafilm®.
Clear
material such as Saran®, widely used in the kitchen for food storage,
should never be used. These materials adsorb large amounts of the dye.
Containers
made of the above materials tend to be more expensive, but are indefinitely
reusable.
An
inexpensive container of adequate size (about 2/3 oz), which we have tested,
is a special low-potassium glass scintillation vial with Polyseal® caps
made by Kimble. They are available from Scientific Products and other
laboratory suppliers.
Filtration
The
requirement is that the sample be free enough of suspended solids so that
accurate measurements of the fluorescence of the sample may be made.
In
most cases, filtration will not be required. Whether filtration is necessary
and adequacy of the filtration procedure may be easily checked. See PROOF
OF PERFORMANCE -- Recovery Ratio Test.
The
Effect of Turbidity
Your
fluorometer shines green light on the dye to be measured, and responds
to the red light which the dye emits. If these two beams of light are
partially obscured, then readings will be low.
Note
that the effect is a proportional one, not a subtractive one. If 10% of
the light is lost due to turbid materials, then all readings will be reduced
by 10%, regardless of whether they are high or low.
Note
also that the turbid material must remove light to affect the readings.
If the particles present do not absorb light, but merely deflect it, readings
will be only slightly affected. It is the big, dark particles that do
the damage.
The
"fines", which are often very difficult to remove will cause
very little error.
Further,
we find that the amount of big, dark particles is quite variable, but
the content of "fines" is low and consistent.
There
are three methods which may be used to correct or remove errors caused
by turbidity:
- Prepare
standards with the water of interest. Assuming turbidity is constant,
this automatically corrects for errors.
- Remove
the suspended solids to the point where the error is negligible. If
you are in a hurry, filter; if you are not in a hurry, let them settle.
- Dilute
the samples with clean water to the point where turbidity is negligible
and multiply the answers by the dilution factor. Don't forget that the
blank and standards should be similarly diluted.
Preparation
of standards with the water of interest is valid at almost any level of
turbidity if the turbidity is constant. Even if 50% of the light is lost,
there will be no error. If, however, the turbidity changes by 10% (relative,
i.e., from 50 to 45% light loss) you now have a 10% error in your results.
There is no way to predict how much light loss can be tolerated without
actually testing the variability in light loss. In a crucial meter calibration,
one might collect several blank samples, prepare standards in them and
compare readings; or, simply remove all solids.
In
routine measurements in sewers, we normally recommend that you accept
up to a 10% loss of light. A 10% (relative) change in turbidity is only
a 1% error. The quick filtration techniques described below, typically
reduce turbidity effects to about 2%. Overnight settling of sewage usually
reduces the effect to near zero.
Use
of Existing Filtration Equipment
If
measurements are to be made in a laboratory equipped for filtration of
samples in reasonable volume, then filter. But there are several things
to avoid:
- Common
filter paper should be avoided. Glass fiber filter discs or filters
such as the Millipore® type are O.K.
- Cross-contamination
between one sample and the next must be avoided. Rinsing with clear
water will not solve the problem, since residual clear water will dilute
your sample and cause errors.
Wash down all the area of the filtration apparatus with a small
amount of your sample. Filter it. Swirl the contents of the vessel
which receives the filtered sewage. Discard.
Then,
filter and save for measurement.
In
extreme cases, you may wish to wash down and discard two or three
times, before saving a sample for measurement.
- Cross-contamination
is easy to check. Wash down with water. Filter some sewage containing
dye and save. Repeat several times, saving each sample.
Then
measure, to see how many wash- downs are required.
A
Quick Field Filtration Technique
We
have developed a rapid filtration technique based on very inexpensive
disposable items, which avoids sample cross- contamination and may be
used under field conditions.
The
equipment which we use is available through Scientific Products. (See
Laboratory Equipment and Supplies in your telephone directory yellow pages).
Equivalent items are available from other major supply houses.
- Disposable
plastic funnels. We have found that Lance, 5.5 cm (2-1/4") polystyrene
funnels, manufactured by Sherwood Medical Industries, Inc., are excellent.
They may be rinsed and reused, but are brittle and subject to cracking.
Polypropylene
funnels, though somewhat more expensive are more durable, and a better
choice, if you wish to rinse and reuse them.
- Pyrex®
brand wool filtering fiber, Corning Catalog #3950 or equivalent. A pound
will be a lifetime supply.
- A
test tube rack for 13 mm x 100 mm culture tubes, which may be readily
immersed in water.
- A
pan which you can place the test tube rack in, deep enough so that when
partially filled with water, cuvettes in place in the test tube rack
will be immersed to about 1-1/2" from their tops.
(This is not part of the filtration set-up, but
will be required for measurement.)
- A
roll of Parafilm. One 4 inch x 125 foot roll is a lifetime supply.
Clear
material such as Saran, widely used in the kitchen for food storage
should never be used. These materials adsorb large amounts of the
dye.
- A
pair of scissors.
- A
piece of rod about 3/32" in diameter, 3" long. A 3/32"
Allen wrench with handle is excellent.
You
will need to experiment a little:
- Working
from the large end, force a tightly wadded piece of glass wool into
the stem of the funnel, using the 3/32" rod.
The piece should be large enough so that some
forcing is required to end up with a plug of packed fiber in the stem,
about 3/8" to 1/2" long. One-eighth inch or more of looser
fiber should extend above the stem.
- Place
a 13 mm x 100 mm cuvette in the test tube rack.
Place the funnel, packed with glass fiber in the
cuvette.
- Pour
about one ounce of water into the funnel, and time how long it takes
to fill the cuvette to the label.
It should take between two and five minutes. Adjust
your packing procedure until the filtration rate is correct.
Be
prepared in advance! Prepare as many filters as needed before handling
dye. Have a similar number of clean, dry 13 mm x 100 mm cuvettes on
hand, labeled to match your sampling schedule.
During
the study, always handle the funnels by the outside, near but not
at the top. Similarly, do not touch the rims of the cuvettes. This
will minimize the chance of any dye on your hands getting into the
samples.
Place
the cuvettes in the test tube rack, staggered so that when the funnels
are in place, the funnels do not overlap.
Pour
the sample in. When the cuvette is filled to the label, remove the
funnel and discard. Then, cover the top of the cuvette with Parafilm.
Peel the backing paper from the Parafilm, keeping fingers off the
protected surface. Press this clean surface firmly onto the top of
the cuvette and fold the excess over the side.
Your
samples are now ready for measurement.
Overnight
Settling
On
overnight settling, most sewage will settle to a clarity not quite as
good as a sub-micron laboratory filtration, but considerably better than
the quick filtration. To be strictly correct, you should prepare your
standard dilutions before allowing the solids to settle. We have never
seen a case in sewage where the solids adsorbed the dye, but it could
happen. If, for example, the solids contained activated charcoal, dye
would be adsorbed.
In
filling your cuvettes from settled samples, gently pour off anything that
has floated to the top, then pour the liquid into clean, dry, labeled
cuvettes. Cover with Parafilm.
Preparation
of Standards
Standards
are required to calibrate the fluorometer precisely in terms of the dye
that you injected. Standards must be prepared for each run, using a small
portion of the dye actually injected and using a representative sample
of the sewage taken just prior to the start of dye injection.
For
repetitive work, i.e., monthly calibration of a large flowmeter, only
one standard and one blank will be required.
When
you are just starting up, more standards will be required. To save repetition,
the preparation of all standards that you might conceivably need is discussed.
As you become familiar with the technique, you can judge what simplifications
can be made.
Equipment
Needed
The
following is a list of equipment we have found useful. If you have a chemist
available, he or she may well suggest alternatives immediately available
to you.
The
equipment which we use was purchased from Scientific Products. Equivalent
items are available from all major laboratory supply houses:
- 1
ea.1000 ml volumetric flask
- 6
ea.100 ml volumetric flasks
- 2
ea.Volumetric flasks of adequate capacity to calibrate your injector
(see DYE INJECTION - Calibrating the injector)
- 1
case(500) scintillation vials with Polyseal caps (see SAMPLE AND STANDARD
STORAGE)
- 1
ea.1000 ml (1 ml) Eppendorf pipettor (or equivalent) and disposable
tips. (They are relatively expensive, but are worth it in ease and rapidity
of use. Under field conditions, they frequently prove to be more accurate
than conventional pipettes.)
While
the 1 ml pipettor and the 100 and 1000 ml volumetric flasks will adequately
handle all situations, there will be borderline cases where, though not
strictly necessary, it would be nice to have other volumes. The pipettors
are available in adjustable versions.
Care
of Glassware
Technically,
all glassware should be thoroughly clean and dry before use. Cleanliness
is, however, a matter of definition. What most chemists would consider
very bad practice, can be tolerated here. Visible dirty spots, water marks,
finger prints, etc., are not going to cause error unless they contain
dye. Your primary interest is to see that all previous dye is removed
and that the equipment delivers or contains the proper volume of water.
Volumetric
flasks are designed to contain a volume. They can be used wet, without
error, providing all dye has been rinsed out. Only water above the calibration
mark will cause error, and this will be trivial unless heavy beading is
involved.
Beading
occurs eventually in all glassware due to oils in the air and in the water.
Occasional washing with a warm, mild detergent usually cures this problem.
Resistant cases can sometimes be cured by rinsing with acetone, followed
by washing with detergent. Severe cases may require soaking in acid cleaning
solution.
After
washing, or after use, all glassware should be rinsed repeatedly with
tap water (at least three or four times after no dye can be seen) followed
by one or two rinsings with distilled water. The latter rinse is a precaution
against beading and encrusting with minerals from hard water, if the glassware
is to air dry. It is not necessary if you have soft tap water or intend
to reuse the equipment immediately.
To
rinse volumetric flasks, fill them about 1/5 full, swirl and rapidly invert
so that the liquid rinses all inside surface while emptying. Rinse the
outside neck under the faucet between internal rinsings.
Cuvettes
should be treated like the flasks, except that they must be dry for use.
If they are not dry, they should be rinsed several times with the solution
to be measured. This is not convenient when you are filtering directly
into the cuvette. Cuvettes should be stored to avoid scratches. Minor
scratches will not affect accuracy, but eventually such treatment will
cause errors.
Dilution
Technique
It's
easy to perform an accurate dilution, but there are several points which
must be carefully observed.
Position
yourself so you have good light, and preferably are facing a light-colored
wall.
Pipettor.
Follow the manufacturer's recommendation with your pipettor. Most have
three positions of the activating button. For these, the general instructions
are (after firmly seating a fresh tip):
- Depress
the activating button to the first stop.
- Lower
the pipettor till the tip is slightly below the surface of the sample
to be pipetted.
- Slowly
release the activating button. Wait a few seconds and withdraw the pipettor.
- Check
that there is no air bubble at the end of the tip.
- Tilt
the volumetric flask, and insert the pipettor such that the tip is well
below any ground portion, and is touching the side.
- Gently
depress the activating button past the first stop to the second. This
dispenses the liquid and blows out the last droplet.
- Remove,
and dispose of the tip by depressing the button to the final (eject)
position.
Volumetric
Flask. Fill the flask to the mark with water, or raw sewage containing
no dye, depending on the step in the dilution procedure and type of standard
being prepared. Use a dropper or clean pipette to add the last bit. If
you overfill slightly, don't worry about it. In a typical volumetric flask,
missing the mark by half an inch leads to an error of only about 0.7%.
If you overfill a lot start all over. For highest accuracy, the bottom
of the meniscus should touch the top of the mark.
Cap
the flask, and holding the cap tightly in place, invert. Swirl the flask
and right it until the air bubble reaches the top. Repeat a total of 10
times.
You
can easily check the precision of your dilution technique. Just make several
identical dilutions and check the repeatability, using the Fluorometer.
See MEASUREMENT - Normal Measurements.
Standards
Required
When
making routine measurements, only one standard is required. This is a
precisely known dilution of the injected dye in raw, freshly-stirred,
or mixed sewage. The standard must be treated in the same manner as your
samples, i.e., allow it to settle or filter it.
For
the first measurement in a new system, you will also need a RECOVERY STANDARD.
This is the same precise dilution made above, but with distilled water.
Actually, this recovery standard is very little additional trouble to
prepare. Until considerable experience is gained, it is good practice
to prepare it routinely.
Preparation
of Standards
Referring
to the section on DYE INJECTION - Dye Concentration, an example was given
where a 1% solution was injected with the intent of obtaining a concentration
of 0.1 ppm in the sewer. Let's continue with this example.
While
there is no theoretical requirement that the standard be 0.1 ppm (it can
be anywhere from 0.01 - 0.5 ppm), errors are reduced if the standard is
near the sample concentration. To prepare 0.1 ppm standard, we need to
dilute a 1% solution (10,000 ppm) to 0.1 ppm, a 100,000-fold dilution
factor., This will be done in two steps: one dilution of 100-fold, and
one of 1000-fold. Since sewage may contain large particles capable of
plugging the tip of the pipettor, it is customary to make only the last
dilution with sewage; all others are with distilled water. Since the last
dilution should be as near pure freshly-stirred or mixed sewage as possible,
it will always be 100-fold or greater. As much as 1% clean water present
will not cause a significant error.
In
this case, it is immaterial whether the 100- fold, or the 1000-fold dilution
is prepared first, though doing the 100-fold first is technically the
more correct.
If
you have an adjustable pipettor and desire to make dilutions which are
not a factor of 10, keep it simple. If, for example, you would ideally
like 38,000:1, make a 40,000:1 dilution. One dilution of 0.25 ml (250
ml) to 100, and one of 1 ml to 100 gets you 40,000.
The
nicety of adjusting the standard concentration to be near that of the
sample was much more important with meter fluorometers than with digital.
With the digital fluorometer it is not normally of any particular advantage.
If you are doing meter calibration, it is desirable to avoid the (otherwise
negligible) error of the fluorometer changing from the high to low range,
or vice versa.
We
cannot emphasize enough the necessity for proper mixing techniques. Even
when dilutions are accurate, if you draw your sample from an inadequately
mixed flask, your results are invalid.
Preparation
of Recovery Standard
Repeat
your last dilution with distilled water instead of sewage.
Insurance
Immediately
place at least 20 ml (about 2/3 fluid ounce) of your injected dye and
of all dilutions in clean, dry, marked sample bottles. Save these and
the remainder of all sewage samples, including the blank sewage till you
have calculated all flows and decided the results are reasonable and consistent.
The ability to recheck your dilutions may well save you from re-running
the study.
Proof
of Performance
Significance
of Errors
It
is important to keep in mind the accuracy required. If your problem is
localizing massive groundwater infiltration, then ±5% is probably adequate.
If, on the other hand, you are calibrating a flowmeter used to decide
the ratio of payments between a group of municipalities supporting a single
sewage plant, then you might require ±1%.
Errors
that may result may be estimated, as you go through this section on proof
of performance. So, keep your objective in mind. It may well reduce test
costs significantly.
Sampling
Errors
These
errors and the steps to take to correct them are discussed in SAMPLING
- When Should Samples Be Taken? and SAMPLING - Is Mixing Adequate?
In
addition, loss of dye by adsorption on the walls of the flow channel is
accounted for when you carry out the tests outlined under SAMPLING - When
Should Samples Be Taken? Such adsorption is limited in its capacity and
is reversible. It will show up as a slower-than-normal build up of dye
concentration after injection has started and a slower-than-normal fall
in dye concentration after injection has stopped. If a clean "plateau"
of dye concentration can be reached, accuracy will not be impaired.
Recovery
Ratio Test
The
physical and chemical properties of your sewage may interfere with the
measurement of the fluorescent dyes. A recovery ratio test is used to
determine the degree of such interference and the effectiveness of the
several simple procedures used to reduce this interference.
How
to make measurements with the Model 10-AU Fluorometer are discussed in
the section entitled MEASUREMENTS.
Details
of how to prepare Standards and Recovery Standards are discussed in the
section entitled PREPARATION OF STANDARDS.
To
determine a Recovery Ratio:
- Determine
the fluorescence due to dye added to distilled water (the Recovery Standard).
- Determine
the fluorescence due to the same dilution of dye in raw sewage (the
Standard).
- Divide
the number from step 2 by the number from step 1. This is the Recovery
Ratio.
In
all cases encountered to date (and expected) this ratio will be less than
one.
If
your recovery ratio is around 0.9, expect accuracy of about ±5%. If it
is around 0.95, expect accuracy of about ±2%. For highest accuracy, all
steps possible to obtain a high recovery ratio should be taken.
The
most general cause of low recovery ratios is the masking effect of suspended
solids in the sewage. To improve the recovery ratio, filtration of the
Recovery Standard (and all unknowns) is called for. For details, refer
to the section entitled FILTRATION.
Effect
of pH
Dye
fluorescence drops off (reversibly), if your sewage is more acid than
pH 4 or more alkaline than pH 10.5. Recovery Ratio will be very low at
extreme pH values.
The
pH is easily checked with a pH meter or indicator paper. If the pH is
low, add a pinch of baking soda. The amount is not critical, since it
cannot take the pH too high. If the pH is high, some dilute acid will
have to be added till the pH is correct. It is probable (not verified)
that monosodium phosphate could be used to correct high pH in the same
way that the baking soda is used for low pH.
Repeat
the Recovery Ratio with pre-treated sewage.
If
recoveries are good, flow measurements of high accuracy may be made.
However,
the pH of all samples, blanks and sewage used for making the standards
must be adjusted.
Unusual
Standards
- Dye
adsorption on particulate material.
If there is a large amount of partially burned
coal, or other similar materials present, they will adsorb dye and
cause low Recovery Ratios.
To
check for this problem, filter the sewage before the dye is added,
in the process of making up your Standard.
Recheck
the Recovery Ratio. If it is significantly better, adsorption on particulate
material is indicated.
Since
dye is truly removed from the system, accurate measurements are not
possible under such conditions.
- Dye
not adsorbed on particulate materials, but filtered sewage appears colored.
Dissolved
colored materials can mask fluorescence, just as particulate material
can, causing low Recovery Ratios.
To
check for this, dilute the sewage used for making the Standard until
an acceptable Recovery Ratio is obtained.
Scale
up the amount of dye injected by this factor. Dilute all samples and
blanks and sewage used for making the Standard by this factor with
distilled water. High precision flow measurements may now be made,
at the cost of one extra dilution per sample.
- Chemical
degradation.
The dye is slowly destroyed by very high concentrations
of residual chlorine, much higher than normally encountered in sewage.
Fresh chlorine, however, destroys the dye rapidly. If the dye goes
past a chlorinator, the chlorinator must be turned off for the duration
of the test.
Presumably,
other strong oxidizing or reducing agents could present a problem.
A
series of Recovery Ratios is called for, one done as quickly as possible,
and others at intervals of several hours.
If
degradation is slow, satisfactory results may be obtained by making
up your Standard at the same time that injection is going on, and
measuring your samples as quickly as possible.
Keep
the Standard and all samples at the same temperature until measurements
can be made, since chemical degradation is sharply temperature sensitive.
Help From the Manufacturer
If
you feel that you have a special problem, call our Technical Service Department.
Measurement
Your
Model 10-AU Fluorometer should be equipped with the 13 mm x 100 mm cuvette
holder and the appropriate filters and lamp from the 10-041 or 10-041R
Rhodamine Optical Kit. (Refer to the Model 10-AU Fluorometer User's Manual
for selection and installation instructions.)
The
following information describes, in general, what needs to be done to
measure samples. Step-by-step instructions using the Model 10-AU Fluorometer
are set forth in "A Procedure for the Turner Designs Model 10- AU
Digital Fluorometer: Flow Measurements in Sanitary Sewers by Dye Dilution."
(Detailed operating instructions for the fluorometer can also be found
in the user's manual.)
Sample
Handling
For
precise measurements, samples and standards must be at the same temperature,
since the fluorescence of rhodamine WT has a temperature coefficient of
-2.6% per degree centigrade.
Equipment
required is described under FILTRATION -- A Quick Field Filtration Technique.
The points that should be remembered are:
- All
samples, standards and blanks should be poured into clean, dry 13 mm
x 100 mm cuvettes. (If you have used the Quick Field Filtration Technique,
they will already be in such cuvettes).
The volume of liquid in the cuvettes should be at least 5 ml.
This will be assured if they are filled about two-thirds or more.
Do not overfill, however, as it makes the cuvettes awkward to handle.
- All
samples, standards and blanks should be in your test tube rack, except
when being measured.
- The
test tube rack should be in a pan of water at room temperature.
The
water level in the pan should be high enough so that it covers the
bottom three- fourths of the cuvettes.
- If
samples were at about room temperature, they should be in the water
bath for at least 15 minutes before measurements start.
If they were at a markedly different temperature,
allow 30 minutes.
- Measurements
should be made rapidly, since the instrument is at a higher temperature
than the room.
If a measurement (or instrument adjustment) cannot
be made within 45 seconds, the sample should be returned to the test
tube rack, so that it will return to the correct temperature.
This
is why duplicate Standards are called for. See Samples Required, below.
- We
recommend a minimum of 5 minutes in the test tube rack between one measurement
and the next on a single sample.
- The
outside of the cuvette MUST BE carefully dried with facial tissue before
it is placed in the instrument for measurement.
- The
fluorometer should be allowed to warm up for at least ten minutes prior
to the start of measurements.
Samples
Required
Experience
may allow you to reduce the number of standards, but for initial studies
you should have:
- Two
cuvettes filled with Standard. See Preparation of Standards.
- Two
cuvettes filled Recovery Standard.
- One
cuvette filled with Blank.
If you are studying Blank variability, you may
have a number of cuvettes filled with Blanks. See DYE INJECTION --
Dye Concentration.
- One
cuvette filled with distilled water.
- Your
various samples, each in its own cuvette.
Recovery
Ratio
Recovery
Ratio Tests are important in predicting the precision of your flow measurements
and guiding you in improving this precision. See PROOF OF PERFORMANCE
-- Recovery Ratio Test, Effect of pH, Unusual Situations and Help from
the Factory.
- Insert
the cuvette containing distilled water (outside of the cuvette MUST
BE dry!) and suppress the reading to zero (blanking procedure in the
Calibration section of the user's manual).
- Insert
the cuvette containing your Recovery Standard. Record the reading on
the digital display.
Note:The number you are reading is "relative" concentration.
Do not fall into the trap of believing that this is the exact concentration.
You set up the instrument with a standard of possibly questionable
accuracy, and definitely at a different temperature. You do not need
to know the actual concentrations of your samples, only the relative
concentrations.
- Insert
the cuvette containing your (sewage) Standard. Record the reading.
- Insert
the cuvette containing your (sewage) Blank. Record the reading.
- Subtract
reading 4 from reading 3.
The net reading obtained in step 5 is the reading
due to the dye present in the sewage. Divide the net reading from
step 5 by the reading from step 2, to get the Recovery Ratio.
Reading
Samples
Remember,
the only thing that you need is the ratio of Standard to sample.
- Insert
the cuvette containing your Blank (sewage). Record the reading.
If
you wish, you may set the sewage Blank to zero. However, it is useful
to know how large the blank is. If it is small compared to the standard
and samples, then you are more confident of the relative accuracy
of the test. If a small blank varies by a few percent, it is negligible.
If a large one varies the same percentage, the error in the answer
will be proportionately larger.
Therefore,
we recomment that you leave the distilled water set to zero. This
way you definitely have a record of the magnitude of the blank; repeats
will always have positive values; and you may average three or more
cuvettes of your blank for greater accuracy. You will always work
with the net reading: sample (or standard) minus the blank.
- Insert
the cuvette containing your Standard. Record its reading.
As
in the case of the blank, if you wish you may read and average three
or more cuvettes of your standard.
- Read
and record all samples.
If any sample is unduly high ("OVER"
on the Model 10-AU Fluorometer), see "High Concentration Measurements,"
following.
Reading
High Concentrations
For
concentrations between 0.5 and 2 ppm, the samples may be read from a calibration
curve. This is not usually recommended as the curve must be prepared by
making at least five dilutions, preferably seven or more (0.5, 0.75, 1.0,
1.25, 1.5, 1.75, and 2.0 ppm). For this, you need to determine what dilution
of your injected dye will give you about 2 ppm, and prepare the necessary
dilutions to define the curve. Under no circumstances should you attempt
to carry the curve beyond 2 ppm. The curve will be so flat that accuracy
will suffer.
Unless
there are a very large number of samples falling slightly above the linear
range, the fastest and most accurate way to handle high-concentration
samples is to make a dilution.
There
is no theoretical upper limit to the concentration which can be handled
by dilution. In fact, in unusual situations (very high color, for example),
you may wish to deliberately run the concentration very high so that the
interference from the color may be diluted out. In one instance, the author
ran the concentration to about 100 ppm, and made 1000-fold dilutions.
The sewer line was opaque with riboflavin, and the 1000-fold dilution
allowed measurement to be made with good accuracy.
It
is best to stick to multiples of 10. Prepare a suitable over-range standard.
If your normal dilution factor for the standard was one million, make
one of 100,000 -- whatever is needed to give you a standard, which when
diluted to the same extent as your samples will yield a standard of the
optimum concentration (about 100 ppb).
Dilute
your samples, the new standard, and your blank with distilled water. If
you are running all samples high because of high turbidity or color, then
prepare a recovery standard in the same manner.
In
the equation for calculating flow, D (the dilution of the standard) is
the amount of dilution of the standard above that of the samples. Thus,
if your standard is diluted 10,000-fold, then standard, blank and samples
are diluted another 100-fold, this 100-fold should not be counted. The
purpose of the "D" is to correct for any dilutions made of the
injected solution that are NOT made in the samples.
Flow
Rate Calculation
The
mathematical equation normally used to calculate flow is:
Q
= q C/c
Where:
Q is the
flow being measured
q is the dye injection rate
C is the concentration of the injected dye
c is the concentration of dye measured
downstream of injection
Since
fluorescence is normally proportional to concentration, a more practical
form of this equation is:
Q
= q x Rst D/r
Where:
Rst is
the reading obtained on your standard
r is the reading obtained on your sample
D is the dilution ratio, used to prepare the standard
The
following discussion of Flow Rate Calculation will be based on equation
2.
Sample
Readings
Sample
readings are all net readings. If (sewage) blank was set to zero, then
the readings obtained for samples are the net readings. If the distilled
water blank was set to zero, subtract the (sewage) blank reading from
all readings to obtain the actual net reading.
Injection
Rate
The
Injection Rate is normally determined by finding the time it takes to
fill a volumetric flask, See DYE INJECTION -- Calibrating the Injector.
Since
volumetric flasks are calibrated in milliliters and elapsed time is most
easily measured in seconds, we have a conversion problem. Two handy numbers
are:
milliliters
x .0002642 = U.S. gallons
milliliters x .00003532 = Cubic feet
As
an example, let's presume that when you tested your dye injector, you
found that it took 158 seconds to fill a 100 ml volumetric flask.
This
means that the flow rate was 100 divided by 158 or 0.631 milliliters per
second, or 37.85 milliliters per minute.
If
you wish your final answer to be in mgd, then the rate of dye introduced
must also be in mgd.
To
convert milliliters per minute to gallons per minute, multiply 37.85 by
.0002642, yielding .01 gallons per minute, or 0.00001440 mgd.
If
you wish your final answer to be in cubic feet per second, then convert
milliliters per second (0.631) to cubic feet per second by multiplying
0.631 by .00003532, yielding .00002287 cubic feet per second.
Standard
Dilution Ratio
You
have already selected a Standard Dilution Ratio. See PREPARATION OF STANDARDS.
In
the example given, it was 100,000 (two 100-fold and one 10-fold: 100 x
100 x 10 = 100,000).
Example:
We
wish to calculate the flow in mgd.
q =00.0000144
mgd (See Injection Rate).
Rst =95.4 (See Reading Samples).
r =25.8 (See Reading Samples).
D =100,000 (See Standard Dilution Ratio).
Q =0.0000144 x 95.4 x 100,000/25.8 = 5.32 mgd, or more reasonably stated,
5.3 mgd.
Of
course, if you are measuring several flows from one injection, the obvious
thing to do is calculate q Rst D, as r is the only variable, from sample
to sample.
Alternative
Operation - Continuous Flow
The
"grab-sample" method described above is not the only mode of
operation using the Model 10-AU Fluorometer.
If
the sewage is not unusually turbid or colored, accuracy may be improved
by operating in a continuous-flow mode.
NOTE:
The Model 10-AU-000 Laboratory Fluorometer cannot perform continuous-flow
studies.
In
this mode, sewage is continuously pumped through a flow cell installed
in the fluorometer, via a submersible pump.
Advantages
- No
cuvette error. Standards, blanks, and samples are all measured with
the same cuvette.
- Temperature
correction can be made automatically by the Model 10-AU (if the optional
temperature-compensation feature is purchased).
- Anomalies,
such as rapidly changing flows, become obvious.
- If
the data are collected with a computer or the optional internal data
logger, they may be exported to a spreadsheet program. This permits
the averaging of very large numbers of points for even better accuracy.
Disadvantages
- Longer
setup time.
- More
complicated equipment. A three- way valve should be used such that the
standard and/or blank may be poured or siphoned through without disconnection.
- Power.
A submersible pump that will pump the sewage to the surface from a manhole
is necessary. This generally implies the availability of 115 VAC.
- Standard
preparation requires more time, as one ideally should have several gallons,
usually achieved by multiple use of a 2- liter volumetric flask (the
largest generally available).
- Not
suitable if turbidity is unusually high. Note: The use of a 10 mm diameter
flow cell permits the handling of situations comparable to those using
the 13 mm x 100 mm cuvettes. The 25 mm Continuous-Flow Cuvette System
should be used only with quite clean water.
|