Preparation of Standards
for Dye Studies
using Rhodamine WT
What is a Standard?
A standard is a known concentration of
the dye you are injecting. It is used to calibrate the Turner Designs
Model 10 Fluorometer to the desired sensitivity. The fluorometer reading
of the standard will be compared with the readings of unknown samples
to obtain their concentrations. A known concentration is made by weighing
or measuring a sample of tracer and precisely diluting it.
In many cases, the standard will be a known
dilution of dye, not a known concentration. For example, in flow
rate measurements you are only interested in how much a stream dilutes
the dye, not what the actual concentration is. Thus, you do not need to
know the concentration of the dye (in parts per billion or other units)
when you calibrate your fluorometer. You need only know the dilution factor
for the standard and the fluorometer reading, which you will then compare
with the fluorometer readings for your unknown samples collected after
dye has been injected into the stream. This will allow you to calculate
the extent to which the stream dilutes the dye.
Note: For more information on dye
studies and whether a known concentration or dilution is required, see
the monographs "A Practical Guide to Flow Measurement" and "Flow
Measurements in Sanitary Sewers by Dye Dilution", available from
Turner Designs.
Units of Measurement
Use whatever units suit your purpose best.
The EPA guidelines on allowable levels will typically be in micrograms
per liter (1). Other studies are described in milligrams per cubic meter
(mg/m3) or parts per billion (ppb). The simplest to use when making dilutions
is grams per gram, which, for all practical purposes, is the same as grams
per milliliter.
The easiest way to record and to think
of these units is in exponential notation. If you are not used to this,
and are not comfortable with it, move your decimal point around. It is
very easy, however, to think in exponential terms:
1.3 grams dissolved in 100 ml, makes a
0.013 g/g solution. In exponential notation this is 1.3 x 10-2. If 1 ml
is diluted to 100 ml (100 ml is 102 ml), then the new concentration is
1.3 x 10-4.
Some conversion factors:
1 gram/liter = 1 part per thousand (ppt)
10-3
1 milliliter/liter = 1 part per thousand
(ppt) 10-3
1 milligram/liter = 1 part per million
(ppm) 10-6
1 microgram/liter = 1 part per billion
(ppb) 10-9
1 milligram/cubic meter = 1 part per billion
(ppb) 10-9
Linearity
Note: Rhodamine WT comes as a 20%
solution in water (meaning it is 20% active ingredient). The EPA
guidelines are in terms of active ingredient. Thus, one microgram of the
20% solution in one liter is 0.2 ppb active ingredient (or 20%
of 1 ppb).
Instrument readings for fluorescent dyes
are proportional to concentration (linear) from the lowest detectable
level up to a certain concentration. Above this concentration, a multipoint
calibration curve may be used to obtain concentrations. Then, at a certain
concentration (somewhere at about five to ten times the upper limit for
the linear range) the curve flattens out and eventually takes a nose-dive.
This critical concentration is a function of the compound and of the path
length of the flow cell or cuvette.
For practical purposes (using the Turner
Designs Model 10 Fluorometer equipped with the 25-mm flow cell or cuvette
holder), Rhodamine WT is linear to 0.5 ppm (500 ppb, 5 x 10-7, or 500
micrograms/liter). In terms of active ingredient (the 20% solution of
Rhodamine WT), it is linear to 0.1 ppm (or 100 ppb, 1 x 10-7, or 100 micrograms/liter).
Active Ingredient or 20% Solution?
Since Rhodamine WT comes as a 20% solution,
a decision should be made at the outset about whether to make the standard
as 100% tracer or in terms of active ingredient. The first impulse
is to do all calculations on the basis of the active ingredient.
However, even to meet the EPA guidelines, you need only keep in mind that
20% of the original solution is active ingredient. For other purposes,
it is immaterial whether you take the 20% into account. If you discharge
20 pounds of Rhodamine WT solution, or inject a 10-fold dilution, it doesnt
matter that the original material was only 20% pure. If you consider the
Rhodamine WT solution to be pure tracer, then all dilutions are relative.
For example, if you have made a 100 parts per billion dilution based on
100% tracer, your final dilution will be 20 parts per billion active
ingredient. Whatever method you choose, be sure to clearly mark your dilutions
as to 100% tracer or active ingredient.
Measuring by Weight or Volume
For flow measurements and other studies
where dye is to be added to a body of water, if you plan to add dye in
pounds, grams, or other weight measure, then your standard must be made
by weighing. If you are adding by volume, then the standard must be made
by volume measurements. The important thing is to make your standard the
same way you make the dye concentration to be injected.
Note, however, that most of the literature
cites the specific gravity of Rhodamine WT as 1.2 (sometimes 1.19). Recent
literature accompanying the dye says 1.15. Thus, one gram equals 1/(sp.gr.)
milliliter. If you add by volume, you could be adding as much as 20% more
dye than if you add by weight. Therefore, if you are concerned with absolute
concentrations (as with an EPA study), you should make an initial 100-fold
dilution by weight (or compensate for the specific gravity). A 100-fold
dilution by weight has a specific gravity of 1.002. For the vast majority
of studies, errors of 0.2% in doing further dilutions, whether by weight
or volume, are acceptable.
If you weigh, it is best to weigh directly
into a volumetric flask. (See discussion of flasks below.) This avoids
the problem of having to rinse whatever vessel is used for weighing. All
of the material must wind up in the flask. Since it will be diluted with
water, the rinsing can be done with water.
If you measure by volume, an accurate method
is to use a large-tip pipette of at least 10-ml capacity and measure dye
into a 1-liter volumetric flask (a 100-fold dilution). Even more accurate
is to measure dye with a 20-ml pipette and dilute to 2 liters. The larger
the pipette, the smaller the surface area with respect to the volume,
hence the smaller the error due to incomplete drainage. The error from
a 10-ml pipette will probably be negligible for most work. It certainly
should be less than 1%.
The most accurate way to measure the tracer
by volume is to fill a 10-ml volumetric flask to the mark, then rinse
the tracer into a 1-liter volumetric flask. You cant rinse a pipette,
as it is made to deliver the stated volume. Volumetric flasks contain
the stated volume, so if you want the stated volume out of the flask,
you must rinse.
Note, however, that there is not much point
in preparing your standards to an accuracy greater than the means by which
the tracer will be added to the system.
The Volumetric Flask
Put your measured sample into the flask,
then add water up to the line. The correct level is when the bottom of
the meniscus touches the calibration line. The more precise Class A flasks
will be accurate to 0.03% for a 1-liter flask, and to 0.16% for a 100-ml
one. Less accurate flasks have a tolerance of twice this.
After adding the dye, mix thoroughly. The
proper way is to invert the flask (hold the cap on!). When the air bubble
has risen, swirl it for a few seconds. Right it; let the bubble rise;
then invert and swirl again. Do this ten times.
To care for a volumetric flask, rinse it
thoroughly with distilled water. Occasionally, wash it with a mild detergent.
To make sure it is clean, put a few milliliters of water in it, swirl,
and pour into a cuvette. Then, compare the fluorometer reading with clean
water. Rinse until they read the same. The flask does not have to be dry
to use it.
The Pipette
There are three choices: a pipettor, a
measuring pipette, or a volumetric pipette.
Pipettor. Advantageous for most
users. It is not necessary to learn to control the leakage while you adjust
to a line, and disposable plastic tips are used. Accuracy is generally
about 1%. There are many brands on the market. They can be purchased in
fixed volumes, for example, 100 microliters (0.1 ml), 1 ml; or adjustable
volumes, 10-100 microliters, 0.1-1 ml.
Pipettors are relatively inexpensive and
easy to use: 1) Push a button until you feel a stop; 2) Insert tip in
solution; 3) Release slowly to draw up the set volume; 4) Push past first
stop to a second one to eject and blow out the last drop. Some pipettors
have a further position that ejects the tip.
Measuring Pipette. Available in
disposable form. The accuracy of the disposable ones, +/- 2%, is no worse
than the nondisposable, and a clean one is always available. When using,
let it drain, then blow the last portion out.
Volumetric Pipette. Like the volumetric
flask, this is very accurate. First, fill the volumetric flask about 1/2
full with water. Then, fill the pipette to the line, and, holding the
pipette vertical, allow it to drain. When it stops dripping, hold it for
about 10 seconds longer, then touch the drop on the end to the surface
of the water (just once, and only briefly). Remove the pipette.
Problems: A pipette must be dry
when you use it, or it must be rinsed with the material you intend to
pipette. Volumetric pipettes are difficult to clean and rinse. If you
need the accuracy, buy quantities of them, so you can use them, then clean
them at leisure (preferably with access to a laboratory pipette washer).
Making Standard Dilutions
For a standard, you need any concentration
no greater than the linear range of your tracer (for the rhodamine dyes,
approximately 100-ppb (0.1-ppm; 100-micrograms/liter) active ingredient.
To obtain this concentration, you will
make serial dilutions. By this, we mean you take your concentrated
solution and make a dilution of it. You mix it thoroughly, then make a
dilution of that, and so on, until the desired concentration is obtained.
We recommend preparing a higher concentration,
i.e., around 0.1-ppm (100-ppb or 10-7) active ingredient. At these
high levels (high for fluorescence), contamination will be less of a problem.
Contamination from dirt or other things is not a problem, but spurious
tracer could be. In preparing the standards, you are handling the pure
material and high concentrations, and it is safer to use the highest standard
that is convenient.
Note that the dilutions you are after can
be achieved in a variety of ways. The easiest way is with 1- and 10-ml
pipettors, and a choice of 100-ml and 1-liter flasks. If intermediate
concentrations are desired, use an adjustable pipettor, or pipette several
shots into one flask, or use intermediate-size flasks (they are available
in 200-, 250-, and 500-ml sizes). Generally, all you are after is some
concentration not greater than 500 ppb (0.5 ppm)--or 100 ppb (0.1 ppm)
if dealing in active ingredient. Since readings are proportional
to concentration at or below this point, it is simply a question of convenience.
Dont use all clean water.
Your last dilution should always be done twice, once in distilled water,
and once in the water in which the measurement will be made. This is because
sometimes there are substances in the test water that interfere with the
reading. This doesnt happen often, but it can invalidate your readings
if you dont recognize it.
Your standard will be the dilution in
the system water, but first you need to see that it reads the same
as the dilution in distilled water -- or make sure you understand any
difference.
To prepare a 100-ppb (active ingredient)
standard of rhodamine WT (20% solution):
- First, prepare a 100-fold dilution by
weight. (See section MEASURING BY WEIGHT OR VOLUME for an explanation.)
Using an accurate laboratory scale, weigh 1 gram of dye directly into
a 100-ml volumetric flask. The dye may be dripped into the flask with
a pipette until 1 gram is obtained. Then dilute to the mark with distilled
water. You now have a 10-g/liter (10 ppt, 10-2) concentration of your
tracer.
Note 1: You could obtain the same concentration
by weighing 10 g into a 1-liter flask, or 20 g into a 2-liter flask.
Note 2: If you intend to inject
dye by volume, then pipette 1 ml of dye into a 100-ml volumetric flask
and dilute to the mark with distilled water. Or measure 10 ml of dye
into a 10-ml volumetric flask and rinse into a 1-liter flask. Then,
dilute to the mark with distilled water. This will yield a 10-ml/liter
(10-ppt, 10-2) dilution. (Keep in mind the specific gravity factor.
See MEASURING BY WEIGHT OR VOLUME, above.)
- Next, pipette 1 ml (or weigh 1 gram)
of the dilution in #1 (10-2 or 10 ppt) into a clean 100-ml volumetric
flask and dilute to the mark with distilled water. Mix thoroughly. You
now have a 10-4, or 100-ppm, dilution.
- Now, pipette 5 ml (or weigh 5 grams)
of the dilution in #2 (10-4 or 100 ppm) into a clean 1-liter volumetric
flask and fill to the mark with system water. Mix thoroughly.
You now have a 10-7 (or 100-ppb; 0.1-ppm) active ingredient standard.
Note: We measured 5 ml because
rhodamine WT comes as a 20% solution (meaning 20% active ingredient).
If you are not concerned with active ingredient, then diluting 1 ml
1000-fold yields a 100-ppb dilution of tracer (or 20-ppb active
ingredient).
- Repeat step 3, using distilled water.
Compare fluorometer readings of this dilution with that of #3.
To prepare a 100-ppb standard of rhodamine
B or other dye in powder form:
- First, prepare a 100-fold dilution by
weight. (See MEASURING BY WEIGHT OR VOLUME above, for an explanation.)
Using an accurate laboratory scale, weigh 1 gram of dye directly into
a 100-ml volumetric flask. Then dilute to the mark with distilled water.
(Be sure to mix thoroughly; the powders can be difficult to mix.) You
now have a 10-g/liter (10-ppt, 10-2) concentration of your tracer.
Note: You could obtain the same concentration
by weighing 10 g into a 1-liter flask, or 20 g into a 2-liter flask.
- Next, pipette 1 ml (or weigh 1 gram)
of the dilution in #1 (10-2 or 10 ppt) into a clean 100-ml volumetric
flask and dilute to the mark with distilled water. Mix thoroughly. You
now have a 10-4, or 100-ppm, dilution.
- Now, pipette 1 ml (or weigh 1 gram)
of the dilution in #2 (10-4 or 100 ppm) into a clean 1-liter volumetric
flask and fill to the mark with system water. Mix thoroughly. You now
have a 10-7 (or 100-ppb; 0.1-ppm) standard.
Note: Rhodamine B is also available in
30% or 40% active ingredient solutions. If you are working
with one of these dilutions, then what you actually have is a 30-ppb
(30%) or 40-ppb (40%) standard, based on active ingredient.
- Repeat step 3, using distilled water.
Compare fluorometer readings of this dilution with that of #3.
Thus, the necessary concentration for your
standard can be achieved in a variety of ways. For example, a 5 x 10-7
or 500-ppb dilution could be achieved in the following cases:
- You are doing a flow measurement, and
have a dye concentration of 5 x 10-3 (5 ppt). To achieve a 5 x 10-7
(500-ppb) dilution, make two 100-fold dilutions of the 5 ppt.
- You have a 5% solution (5 x 10-2) of
dye; make one 1000-fold dilution and one 100-fold dilution; or a 10-fold,
followed by two 100-fold dilutions.
Making more than one Concentration
Whether or not you make more than one concentration
depends on such considerations as the confidence you have in your dilution,
and how important it is that your standard be exactly right. If this is
your first time, it would be important to prepare a range of concentrations
and plot a standard curve. This will test your proficiency and make you
more confident. The readings, however, should be linear with concentration.
You really need only one concentration (and a blank).
The most important thing to do is to be
sure there were no errors, i.e., a lapse in counting measurements, an
accidental contamination, an air bubble in the pipetted sample, etc.
It is not so much, therefore, the need
for more than one concentration as it is a need to duplicate your preparation.
This means from the beginning. If you choose to make several concentrations
at the final dilution, fine.
pH, Turbidity, and Chlorine
pH. When you are taking measurements,
the most serious, nonvisible problem in test waters is pH. Any pH between
4.5 and 10.5 is fine. Most systems should fall in this range, but if the
pH does not, the fluorescence will drop off rapidly. It is, however, reversible.
If the dilution in the system water reads very low and there is no obvious
reason (intense color, very high turbidity, etc.), check the pH.
If you cant check the pH, get some vinegar and some baking soda.
Try adding a pinch of the soda to one test tube, and a drop or two of
vinegar to another. Neither is capable of taking the pH too far in the
other direction. If this causes the reading to increase, add a bit more
to see if you have enough, then plan on adding the same amount to all
your samples.
If pH is not the problem, then the study
probably cannot be done with a fluorescent tracer. This is extremely rare,
and it is likely that someone is dumping a high concentration of a very
strong oxidizing agent. Investigate.
Turbidity and Color. These are covered
in the monograph "A Practical Guide to Flow Measurement. Generally,
it takes considerable turbidity or color to interfere with the readings.
The interference is a percentage reduction in reading. For example,
say 100 ppb reads 900 in clean water. A 10-ppb solution would read 90.0,
and a 1-ppb solution, 9.00. Your 100-ppb solution in the system water
reads 810 (a 10% reduction). The 10-ppb solution will read 81.0 and the
1-ppb solution, 8.10. In other words, if you use the dilution in the system
water as the standard, there will be no error, and no correction needs
to be made.
Should you calibrate this way if there
is a 90% loss of reading? Theoretically you could, if you were absolutely
positive that the turbidity or color would be constant during the study.
In practice, it would be much better to increase the dye concentration
by a factor of 10, then dilute all samples, the blank, and the standard
10-fold with clean water.
pH and Chlorine. In potable water,
we found in lab tests that chlorine appears to destroy rhodamine WT within
a few minutes at all pH levels, even with very low levels of chlorine
(.1 part per million). There is at least one very thorough published study
that showed little effect of chlorine on rhodamine WT in wastewater (2).
We speculate that the suspended solids in wastewater have a prophylactic
effect.
Storing Samples and Standards
The tracers will not degrade, and if stored
in the dark, are stable for years. After you have completed your study,
you may find that something doesnt fit. If you have a sample of
the tracer you injected, samples of your dilutions, and your field samples,
you can always re-read your dilutions and see what happened. Since the
Model 10 Fluorometer 10-030 Cuvette System requires only 4 ml of sample,
there is no need to store large amounts. Scintillation vials (discussed
in Flow Measurements in Sanitary Sewers) hold about 20 ml,
and 500 of them require very little storage space.
Sources of Dyes
Rhodamine WT and Rhodamine B -- Industrial
grade of D & C Red #19
CROMPTON & KNOWLES CORPORATION
P.O. Box 33157
Charlotte, NC 28233-3157
1 (800) 432-6188
FORMULABS
1710 Commerce Drive
Piqua, OH 45356
(513) 773-0600
KEYSTONE CORPORATION - Chicago Division
2501 W. Fulton Street
Chicago, IL 60612
1-800-522-4DYE (4393)
Website: www.dyes.com
KEYSTONE CORPORATION - Pacific Division
13767 Milroy Place
Santa Fe Springs, CA 90670
1-800-447-4DYE (4393)
Website: www.dyes.com
Fluorescien (also referred to as
Uranine)
PYLAM PRODUCTS COMPANY, INC.
100 Stewart Ave.
Garden City, NY 11530
(516) 222-1750
(800) 645-6096
KEYSTONE CORPORATION - Chicago Division
2501 W. Fulton Street
Chicago, IL 60612
1-800-522-4DYE (4393)
Website: www.dyes.com
KEYSTONE CORPORATION - Pacific Division
13767 Milroy Place
Santa Fe Springs, CA 90670
1-800-477-4DYE (4393)
Website: www.dyes.com
Sulpho Rhodamine B Extra
PYLAM PRODUCTS COMPANY, INC.
100 Stewart Ave.
Garden City, NY 11530
(516) 222-1750
(800) 645-6096
References
- U.S. Environmental Protection Agency
(Office of Drinking Water), "Guidance Manual for Compliance with
the Filtration and Disinfection Requirements for Public Water Systems
Using Surface Water Sources," Report No. EPA 570/9-89-018, Appendix
C-9 (October 1989).
- (0026) D.G. Deaner, "Effect of
Chlorine on Fluorescent Dyes," J. of Water Poll. Control Fed.,
45:3, 507-514 (1973).
|